Safety engineered injection devices for intramuscular, subcutaneous and intradermal injections in healthcare delivery settings: a systematic review and meta-analysis

verfasst von: Alain C. Harb, Rami Tarabay, Batoul Diab, Rami A. Ballout, Selma Khamassi, Elie A. Akl

Erschienen in: BMC Nursing | Ausgabe 1/2015

Abstract

Background

Occupational sharps injuries are associated with transmission of bloodborne viruses to healthcare workers, including hepatitis B virus (HBV), hepatitis C virus (HCV), and human immunodeficiency virus (HIV). Similarly reuse of syringes in healthcare settings might transmit these infections between patients. The objective of this study was to systematically review the evidence about the effects of the use by health care workers of two types of safety engineered injection devices, when delivering intramuscular, subcutaneous, or intradermal injectable medications: sharps injury protection syringes and reuse prevention syringes.

Methods

We included both randomized and non-randomized studies comparing safety syringes to syringes without safety features. Outcomes of interest included needlestick injuries, and HIV, HBV and HCV infections amongst HCWs (for sharps injury prevention syringes) and patients (for reuse prevention syringes). When possible, we conducted meta-analyses using a random-effects model. We tested results for heterogeneity across studies using the I statistic. We assessed the quality of evidence by outcome using the GRADE methodology.

Results

We included nine eligible studies: six assessed devices that qualify as sharps injury prevention devices, and three assessed devices that qualify as both injury prevention devices and reuse prevention devices. Eight studies were observational while one was randomized. All studies assessed a single outcome: needle stick injuries among healthcare workers. For sharp injury prevention syringes, the meta-analysis of five studies resulted in a pooled relative risk of 0.54 [0.41, 0.71] for the effect on needlestick injuries per healthcare worker. The associated quality of evidence was rated as moderate. For reuse prevention syringes, data from one study provided a relative risk of 0.40 [0.27, 0.59] for the effect on needlestick injuries per healthcare worker. The associated quality of evidence was rated as moderate. We identified no studies reporting on the effect on the reuse of syringes.

Conclusions

We identified moderate quality evidence that syringes with sharps injury prevention feature reduce the incidence of needlestick injuries per healthcare worker. We identified no studies reporting data for the remaining outcomes of interest for HCWs. Similarly we identified no studies reporting on the effect of syringes with a reuse prevention feature on the reuse of syringes or on the other outcomes of interest for patients.
Hinweise

Electronic supplementary material

The online version of this article (doi:10.​1186/​s12912-015-0119-1) contains supplementary material, which is available to authorized users.

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

Conceiving and designing the review: AH, RT, SK, EA. Coordinating the review: EA. Data extraction: AH, RT, RB, BD. Data analyses: EA. Data interpretation: SK, EA. Writing of the review: AH, EA. Reviewed and approved the final version of the manuscript: all authors.
Abkürzungen
CDC
Centers for Disease Control
DALY
Disability-adjusted life years
HBV
Hepatitis B virus
HCV
Hepatitis C virus
HCW
Healthcare workers
HIV
Human immunodeficiency virus
ID
Intradermal
IM
Intramuscular
NSI
Needlestick injuries
RR
Relative risk
SC
Subcutaneous
WHO
World Health Organization

Background

Healthcare workers (HCWs) exposure to bloodborne pathogens from sharps injuries, primarily needlesticks, is a serious occupational problem. The World Health Organization (WHO) reported that more than three million HCWs were exposed to bloodborne pathogens from percutaneous exposure in the year 2000 across the world [1]. In the United States alone, and according to the Centers for Disease Control and Prevention (CDC), hospital-based HCWs suffer about 385,000 such injuries annually. This amounts to an average of 1000 injuries per day [2]. In the United Kingdom, sharps injuries account for 17 % of accidents to the National Health Services staff [3].
Occupational sharps injuries are associated with transmission of bloodborne viruses, the most serious and potentially fatal of which are hepatitis B virus (HBV), hepatitis C virus (HCV), and human immunodeficiency virus (HIV) [4, 5]. They are also associated in the transmission of more than 20 other pathogens [4, 5]. It has been estimated that occupational sharps injuries are responsible for 32 % of HBV infections, 40 % of HCV infections, and 5 % of HIV infections [6]. The HCWs risk of sharp injury related infection is relatively high in Africa, where HIV is prevalent and HBV is endemic amongst the patient population [7].
The burden of sharp injuries affects both HCWs and healthcare institutions. Sharp injury related blood-borne infections lead to absenteeism, morbidity and, mortality among HCWs [8]. They may also induce psychological stress, and negatively affect the personal and work life of HCWs [9, 10]. Hospitals also suffer from costs related to testing, treatment, and lost working time [11].
Reuse of syringes in healthcare settings can transmit these infections between patients. In the year 2000, the reuse of injection equipment accounted for 32, 40 and 5 % of new HBV, HCV, and HIV infections worldwide [12]. The estimated burden related to this practice is around 9.18 million disability-adjusted life years (DALYs) between the years 2000 and 2030 [13].
One of the suggested interventions to reduce sharps related injuries is the use of safety-engineered devices, which have mechanisms to prevent percutaneous injuries [14]. Indeed, introducing the use of these devices may prevent sharp injuries and the associated bloodborne infections [15]. Safety features of safety-engineered devices are designed to shield the needle or non-needle sharp object after use. There are also two main types of safety syringes:
(1)
Sharps injury prevention syringes: these use different mechanisms e.g. self-retractable needles, internal blunt needles, or external shielding
 
(2)
Reuse prevention– syringes: these include a reuse prevention feature e.g. metal clip to block the plunger once the injection is given, the plunger breaks etc. (making them unusable after initial use).
 
We conducted this study in preparation for the development of WHO policy guidance on use of safety-engineered devices by healthcare workers to deliver IM, SC and ID injections. The objective was to systematically review the evidence about the effects of the use by health care workers of two types of safety devices: sharps injury prevention syringes and reuse prevention syringes.
The specific review questions were:
1.
What are the benefits and harms of sharps injury prevention syringes versus single use disposable syringes when used by healthcare workers to deliver intramuscular, subcutaneous or intradermal injections to patients?
 
2.
What are the benefits and harms of reuse prevention syringes versus single use disposable syringes when used by healthcare workers to deliver intramuscular, subcutaneous or intradermal injections to patients?
 

Methods

The study consisted of a review of the literature and did not involve any ‘human subjects’.

Protocol and registration

We developed two separate protocols for sharp injury prevention syringes and reuse prevention syringes. We registered the protocols with the International database of prospectively registered systematic reviews in health and social care (PROSPERO) [10, 11].

Eligibility criteria

Types of studies included

We included both randomized trials and non-randomized studies including:
  • Cohort studies
  • Case control studies
  • Before and after studies
  • Time-series analysis
We excluded scientific meeting abstracts, research letters, qualitative studies, letters to the editor, reviews, case reports, and case series.

Types of participants and settings

We included studies of healthcare workers delivering intramuscular, subcutaneous, or intradermal injectable medications. We were not interested in non-healthcare settings (e.g., illicit drug use, patients using insulin pen needles). We were not interested in other types of injections (e.g., phlebotomy or intravenous, articular, intra cardiac, and intra peritoneal injections).

Types of interventions

We included studies assessing the introduction of a safety device (sharp injury prevention syringes or reuse prevention syringes) into a healthcare setting. This introduction could have been accompanied by training of HCWs. Eligible sharp injury prevention syringes included: retractable syringes; needle shields, and recapping devices; needleless injectors; needle-safety devices; Eligible reuse prevention syringes included: auto-disable syringes (earlier called “auto-destruct syringes”) (ISO 7886–3), typically meant for vaccination; reuse prevention devices for therapeutic injections (ISO 7886–4); and pre-filled syringe with reuse prevention feature.
Ineligible devices included: intravenous devices; needless adaptors; fistula needle; IV catheters; winged steel needle; implantable port needles; suture needles; all blood collection devices (lancet devices, vacuum tubes for blood collection devices, an arterial blood syringes).
We included studies assessing the introduction of both eligible and ineligible devices as long as they reported data for eligible devices separately. We included the studies not reporting data for eligible devices separately in a sensitivity analysis.

Types of comparison(s)

We included studies comparing one of the interventions of interest to using a device without a safety feature, such as the ‘single use disposable syringes’ (ISO 7886–1).

Outcomes

We included studies assessing at least one of the following outcomes for sharps injury prevention syringes:
  • HIV, HBV, and HCV infections amongst HCWs
  • Other blood-borne infections (e.g. viral hemorrhagic fevers) amongst HCWs
  • Abscesses (septic, aseptic) amongst HCWs
  • Needlestick injuries amongst HCWs
  • Quality of life amongst HCWs
  • Social impact (e.g., stigma, job loss) amongst HCWs
We included studies assessing at least one of the following outcomes for reuse prevention syringes:
  • Reuse of syringes
  • HIV, HBV, and HCV infections amongst patients.
  • Other blood-borne infections (e.g. viral hemorrhagic fevers) amongst patients
  • Quality of life amongst patients
  • Social impact (e.g., stigma, loss of job) amongst patients
  • Needlestick injuries, HIV, HBV, and HCV infections amongst HCWs
Any positive impact on those outcomes would be considered as a benefit, while any negative impact on these same outcomes would be considered as harm.
We used the OVID interface to electronically search in October 2013 the following databases, starting with the dates of their inception: MEDLINE, EMBASE, and CINAHL. OVID is a platform that provides access to online bibliographic databases, academic journals, and other products, chiefly in the area of health sciences. We also electronically searched in October 2013 the Cochrane Central Register of Controlled Trials (CENTRAL). We did not use any study design filter, as we wanted to capture different types of study designs, particularly both randomized and non-randomized studies. We did not use language or date restrictions. Additional file 1 lists the search strategies used. We removed duplicates using the ‘find duplicates’ function in the EndNote software. In addition to the search of electronic databases, we reviewed the references lists of relevant papers; contacted experts; and searched personal files for both published and unpublished studies.

Selection process

The reviewers were organized into two teams of two. Prior to starting the selection process, we conducted calibration exercises to clarify the eligibility criteria. We reviewed 100 citations with every exercise. We achieved agreement by the third exercise. At that point, the two review teams started screening titles and abstracts of identified citations in duplicate and independently. We obtained the full texts for citations judged as potentially eligible by at least one reviewer. Then, the two review teams screened the full texts in duplicate and independently for eligibility. They used a standardized and pilot tested full text screening form. The reviewers compared results and resolved disagreements by discussion or with the help of a third reviewer. We calculated agreement between reviewers for full text screening using the kappa statistic.

Data abstraction process

The two review teams abstracted data from eligible studies in duplicate and independently. They used a standardized and pilot tested data abstraction form with detailed instructions. Then, the reviewers compared results and resolved disagreements by discussion or with the help of a third reviewer. The data items abstracted included:
  • Description of the study device
  • Study design
  • Characteristics of participants and setting
  • Description of the intervention
  • Description of the control
  • Outcomes assessed and statistical results
  • Funding and disclosed conflicts of interest

Risk of bias assessment

The two review teams assessed the risk of bias in each study in duplicate and independently. They used a standardized and pilot tested data abstraction form with detailed instructions. Then the reviewers compared results and resolved disagreements by discussion or with the help of a third reviewer. According to recommendations outlined in the Cochrane Handbook, we used the following criteria for assessing the risk of bias in randomized studies:
  • Inadequate sequence generation;
  • Inadequate allocation concealment;
  • Lack of blinding of participants, providers, data collectors, outcome adjudicators, and data analysts
  • Incompleteness of outcome data;
  • Selective outcome reporting, and other bias.
We used the following criteria for assessing the risk of bias in non-randomized studies:
  • Failure to develop and apply appropriate eligibility criteria
  • Flawed measurement of exposure
  • Flawed measurement of outcome
  • Failure to adequately control confounding
  • Incomplete follow-up
We judged each potential source of bias as high, low or unclear risk of bias.

Data synthesis

For categorical data, we calculated the relative risk (RR) for each outcome for each study. RR refers to the risk in the intervention group or period (e.g., the introduction of a safety device) relative to the risk in the control group or period (e.g., using a device without a safety feature).
We assumed that variability in the population, interventions, control, and outcome measurements across studies will introduce heterogeneity in findings across those studies. To minimize this heterogeneity, we analyzed separately data for sharps injury prevention syringes and data for reuse prevention syringes. Also we analyzed separately different measurement of the same outcome, e.g., NSI per device and NSI per HCW. In order to deal with residual heterogeneity, we then pooled the results of studies using Mantel-Haenszel method (with hybrid inverse variance weighting) to accommodate random-effects. We did not choose the fixed-effects model because it assumes a common effect size, and it is inaccurate with a very small number of studies [16, 17]. In a random-effects meta-analysis the treatment effects for the individual studies are assumed to vary around some overall average treatment effect.
We tested results for heterogeneity across studies using the I statistic. We considered heterogeneity to be substantial if I is greater than 50 %. We planned to create inverted funnel plots of individual study results plotted against sample size in order to check for possible publication bias.

Sensitivity analysis

We identified two studies that assessed both devices for intravenous injections or phlebotomy and devices for intramuscular, subcutaneous or intradermal injections, without providing data separately for the different types of devices [18, 19]. In a post hoc decision, we included these studies in the main analysis but excluded them in a sensitivity analysis to test the impact of their data on the final results. We used the freely available software RevMan 5.1.0 for all analyses [20].

Subgroup analysis

We planned to explain heterogeneity, if present, by conducting subgroup analyses based on the following factors: route of injection (intramuscular, intradermal, subcutaneous), the type of device, level of expertise of HCWs, and time of injury (before, during, or after the injection). In order to assess the effects of reuse prevention devices, and given we did not identify any study assessing a device that is purely a reuse prevention device, we conducted a subgroup analyses of studies of devices that qualified as both reuse prevention devices and sharps injury prevention devices [15, 21, 22].

Quality of evidence assessment

We assessed the quality of evidence by outcome using the GRADE methodology [23].
We produced a GRADE Evidence Profile to summarize the statistical findings and quality of evidence by outcome.

Results

Study selection

Figure 1 shows the study flow. Out of a total of 6566 identified citations, we judged nine as eligible for this systematic review [15, 18, 19, 21, 22, 2427]. Agreement between reviewers for full text screening was high (kappa statistic = 1). Additional file 2 provides the list of the 32 excluded studies with the following reasons for exclusion: reporting on preferences, acceptability or feasibility (n = 5); reporting economic analysis (n = 4); evaluating glucometer lancets (n = 1); reporting data not in healthcare setting (n = 1); and evaluating intravenous injection or phlebotomy safety devices (n = 22) [10, 2857].

Study characteristics

Additional file 3 provides the list of the nine included studies with detailed description of their characteristics.

Type of injection

Out of the nine included studies, five assessed devices for intravenous injection or phlebotomy, in addition to intramuscular, subcutaneous or intradermal injection devices [15, 18, 19, 24, 27]. Of these five studies, three provided data separately for the different types of devices [15, 24, 27]. The remaining two studies reported data combined for the different type of devices [18, 19].

Types of devices

Out of these nine studies, six assessed devices that qualify as sharps injury prevention devices [18, 19, 2427], while three assessed devices that qualify as both sharps injury prevention devices and reuse prevention devices [15, 21, 22].
No studies included a comparison between active and passive devices. Two studies reported SIDs with active safety features [22, 26], two studies reported SIDs with passive safety features [19, 21] and 5 studies had SIDs with both active and passive or unspecified safety features [15, 18, 24, 25, 58].

Device brand

Six specified the device brand and/or the manufacturer:
  • Monoject™ Safety Syringe by Sherwood Medical [21]
  • VanishPoint® by Retractable Technologies, Inc. [15, 22]
  • SafetyGlide™ devices [19], Eclipse™ [25, 27] by Becton Dickinson
  • Surshield™ device by Terumo [27]
The remaining three studies specified neither the brand nor the manufacturer [18, 24, 26].

Funding

Five studies reported their funding sources as follows:
  • Sherwood Medical; [21]
  • Becton Dickinson; [19]
  • National Institute of Allergy and Infectious Diseases; the Centers for Disease Control and Prevention; and the Prevention Epicenters; [24]
  • Directorate General of Public Health of the Autonomous Community of Valencia, Spain; [27]
  • Dutch Ministry of Social Affairs and Employment support; [25]
The remaining four studies did not report their funding sources [15, 18, 22, 26]. Two of these studies evaluated VanishPoint® by Retractable Technologies, Inc. [15, 22] while the other two specified neither the brand nor the manufacturer of the device under evaluation [18, 26].

Conflicts of interest

Two studies reported that their authors had no conflicts of interest [26, 27] The remaining studies did not provide conflicts of interest disclosures.

Study design

One study was a cluster prospective randomized controlled trial [25]. The remaining eight studies were non-randomized and used a before and after study design. Out of these 8 studies, five collected data prospectively [15, 19, 21, 24, 27] while three collected the data retrospectively [18, 22, 26].

Settings

Included studies were all conducted in the following high income countries: Australia (n = 1); [15] Germany (n = 1); [26] Netherlands (n = 1); [25] Spain (n = 1); [27] United Kingdom (n = 1); [19] and United States (n = 4) [18, 21, 22, 24].

Intervention

Interventions consisted of the introduction of the safety devices detailed above under “Device brand”. For seven studies, it was reported that healthcare workers received some form of educational intervention with regards to using the safety devices [19, 21, 22, 2427].

Control

All included studies reported using “standard”, “conventional” or “traditional” syringes in the ‘before’ phase. One study reported conducting a needle safety workshop in the control group [25].

Outcomes

All studies assessed needle stick injuries among healthcare workers. None of the studies reported valuable data on any of the other outcomes of interest. Whitby et al. reported the following: “No significant increase in bloodstream infections was detected during the study period” [15].

Risk of bias within studies

Additional file 4 details and Fig. 2 summarizes the risk of bias in the included randomized study. The trial was at high risk of bias in relation to four out of 7 criteria assessed.
Additional file 5 details and Figs. 3 and 4 summarize the risk of bias in the included non-randomized studies. While the non-randomized studies were generally at low risk for bias in relation to the appropriateness of eligibility criteria, measurement of the intervention, and measurement of the outcome, they were all at unclear risk of bias in relation to dealing with confounding and completeness of data.

Meta-analyses for sharps injury prevention syringes

Eligible studies reporting on the needlestick injuries (NSI) used two main types of statistics: incidence of NSI per device used (or purchased) [19, 21, 22, 27], and incidence of NSI per healthcare worker [15, 18, 2426]. We conducted separate meta-analyses for these different statistics. One study reported incidence of NSI per patient [27].
NSIs reported for all studies were converted to a per year basis.

Needlestick injuries

NSI per device

The meta-analysis of four studies resulted in a pooled relative risk of 0.08 [95 % Confidence Interval (CI) 0.02, 0.27] (Fig. 5) [19, 21, 22, 27]. The I value was 51 %. The inverted funnel plot, although based on only five studies, did not suggest any publication bias (Fig. 6). The sensitivity analysis excluding the one study that did not report separately data for devices for intramuscular, subcutaneous or intradermal injection devices, [19] resulted in a pooled relative risk of 0.12 [95 % CI 0.02, 0.75] and I value of 50 %.

NSI per healthcare worker

The meta-analysis of five studies resulted in a pooled relative risk of 0.54 [0.41, 0.71] (Fig. 7) [15, 18, 2426]. The I value was 43 %. The quality of evidence was rated as moderate (Table 1). Of note, one of the included studies reported data based on a time-series analysis, but we opted to analyze it as a before and after study in order to include it in the meta-analysis. The sensitivity analysis excluding the one study that did not report separately data for devices for intramuscular, subcutaneous or intradermal injection devices [18], resulted in a pooled relative risk of 0.53 [0.36, 0.79] and I value of 56 %. Restricting the analysis to the only included randomized trial resulted in a relative risk of 0.72 [0.30, 1.77].
Table 1
Evidence profile for sharp injury prevention syringes versus single use disposable syringes when used by healthcare workers to deliver intramuscular, subcutaneous or intradermal injections
Quality assessment
№ of patients
Effect
Quality
Importance
№ of studies
Study design
Risk of bias
Inconsistency
Indirectness
Imprecision
Other considerations
IM/SC/I Injury prevention safety devices
No safety devices
Relative (95 % CI)
Absolute (95 % CI)
HIV, HBV and HCV infections amongst HCWs: not reported
            
Critical
NSI per HCW
6
Observational
Not serious
Not serious
Not serious
Not serious
Large effect size
168/16151 (1.0 %)
273/14888 (1.8 %)
RR 0.53 (0.41 to 0.69)
9 fewer per 1000 (from 6 fewer to 11 fewer)
Moderate
Important

Other outcomes

None of the included studies reported data for the remaining outcomes of interest for sharps injury prevention syringes. Whitby et al. made the following statement without reporting any statistical data: “No significant increase in bloodstream infections was detected during the study period” [15].

Meta-analyses for reuse prevention syringes

As stated earlier, three studies reported on devices that qualified as both reuse prevention devices and injury protection devices [15, 21, 22]. Therefore, we conducted subgroup analyses of those studies.

Needlestick injuries

NSI per device: The meta-analysis of two studies resulted in a pooled relative risk of 0.07 [0.01, 0.43] (Fig. 8) [21, 22]. The I value was 41 %.
NSI per healthcare worker: Data from one eligible study indicate a RR of 0.40 [0.27, 0.59] (Fig. 9) [15]. The quality of evidence was rated as moderate (Table 2).
Table 2
Evidence profile for sharp reuse prevention syringes versus single use disposable syringes when used by healthcare workers to deliver intramuscular, subcutaneous or intradermal injections
Quality assessment
№ of patients
Effect
Quality
Importance
№ of studies
Study design
Risk of bias
Inconsistency
Indirectness
Imprecision
Other considerations
IM/SC/I Reuse prevention safety devices
no safety devices
Relative (95 % CI)
Absolute (95 % CI)
HIV, HBV and HCV infections amongst patients: not reported
            
Critical
Reuse of syringes: not reported
            
Important
NSI per HCW
1
Observational
Not serious
Not serious
Not serious
Not serious
Large effect size
35/3253 (1.1)%
77/2829 (2.7)%
RR 0.4 (0.27 to 0.59)
16 fewer per 1000 (from 11 fewer to 20 fewer)
Moderate
Important

Other outcomes

None of the included studies reported data for the remaining outcomes of interest for reuse prevention syringes. As mentioned above, Whitby et al. made the following statement without reporting any statistical data: “No significant increase in bloodstream infections was detected during the study period” [15].

Additional analyses

Although we planned to conduct subgroup analyses to explain heterogeneity, we did not have the opportunity to conduct them mainly because of the relatively small number of studies per analysis. Another reason is the lack of reported data on some of the factors based on which we planned to conduct the analyses: the type of device, level of expertise of HCWs, and time of injury (before, during, or after the injection).
Also, studies did not consistently report stratified outcome data by route of injection (intramuscular, intradermal, subcutaneous).

Discussion

In summary, we identified moderate quality evidence that sharp injury prevention syringes reduce the incidence of needlestick injuries per healthcare worker. We identified no studies, meeting eligibility criteria for inclusion and reporting data for: HIV, HBV, and HCV infections amongst healthcare workers; nor studies, meeting eligibility criteria for inclsion and reporting on the effect of reuse prevention syringes on the reuse of syringes; nor HIV, HBV, and HCV infections amongst patients.
The main limitation in the literature is the lack of evaluation of the effects of the safety devices on outcomes other than needlestick injuries, whether benefits or harms.
Particularly relevant outcomes include the reuse of syringes, or blood borne infections, particularly HIV, HBV, and HCV amongst healthcare workers or patients.
Another limitation related to meta-analytical techniques, is that heterogeneity may be underestimated especially when analyzing a small number of studies [59]. That is why we opted to use the random effect model irrespective of the value of I statistic. Also, given the included studies are relatively old [60], it is likely that publication bias exists and we were underpowered to detect it. Finally, one has to consider that the observed decrease in needle stick injuries shown by the before-after studies, may reflect time trends related to factors such as changes in the legislation, hospital policies, standards of reporting of needle stick injuries.
We have identified two other systematic reviews addressing questions that are similar but not the same as our question [8, 61]. A Cochrane review addressed different types of safety devices for preventing percutaneous exposure injuries caused by needles in healthcare personnel [8]. They found “no clear evidence that the introduction of safe injection devices changed the NSI rate”. In fact, the Cochrane review included only four studies potentially relevant to our review (i.e., injection devices) [18, 25, 27, 62].
While we included three of these studies [18, 25, 27], we excluded the fourth because it was conducted in an educational setting, as opposed to a healthcare delivery setting [62]. Moreover, they analysed two of those studies separately because they reported on multiple safety devices [18, 27]. In our review, we abstracted from those two studies data specific to injection devices and included them in the meta-analysis.
In addition, we included six additional studies not identified by the Cochrane review.
Indeed, the differences in rating the quality of evidence between the Cochrane review and our review could be explained by the differences in study inclusion and the challenges with inter-rater reliability of assessing the quality of evidence [63].
Another review published by the Health and Safety Laboratory for the Health and Safety Executive 2012, addressed different types of safety devices for preventing percutaneous exposure injuries caused by needles in healthcare personnel [61]. They found “there was sufficient published evidence to consider the use of safer sharps devices to reduce the incidence of sharps injuries amongst UK healthcare workers”. In fact, this review included only seven studies potentially relevant to our review (i.e., injection devices) [10, 18, 27, 29, 39, 44, 50]. While we included three of these studies [10, 18, 27], we excluded the other four [29, 39, 44, 50] because these were evaluating intravenous injection or phlebotomy safety devices. Furthermore, they included seventeen other studies that we judged as not eligible for our review.
There is paucity of data about the cost or cost-effectiveness of introducing those devices into healthcare settings. Valls et al. reported that the introduction of sheathed needles for subcutaneous and intramuscular drug administration led to the following changes in cost: −0.010 on hospital wards per patient-day and 0.021 in the emergency department per patient [58]. Whitby et al. reported $46,000 increase in the annual budget of the hospital upon introduction of retractable syringes [15].

Conclusions

The findings of this study have important implications for HCWs practice. Indeed, the introduction of sharps injury prevention devices into healthcare settings is likely to reduce needlestick injuries. Healthcare managers planning to introduce those devices need to consider the cost related to their introduction. They also should do that as part of a comprehensive injection safety program. Such program would include education about the risks associated with accidental injuries, training in using the safety devices, surveillance and reporting of needle stick injuries among HCWs, monitoring and evaluation of the program implementation, immunization of healthcare workers against HBV, and post exposure prophylaxis. In addition, administrators should involve healthcare workers in selecting the devices. Given the paucity of data on the effectiveness of reuse prevention syringes, healthcare managers need to consider their use mainly in settings with high rates of syringe reuse and high prevalence of blood borne pathogens.
The findings of this study have also important research implications. Future studies should assess the impact of introducing safety devices on reuse rates, and on incidence of blood borne infections amongst healthcare workers and patients. In terms of methodology, randomized trials with standardized methods for measuring incidence of sharps injuries would provide better quality evidence relative to currently available evidence. There is also a need to conduct cost-effectiveness studies for different settings, particularly low and middle income countries.

Acknowledgements

We would also like to thank the Injection Safety Program in the Service Delivery and Safety Department of the WHO HQ in Geneva for their financial support. We would like to extend our gratitude to Mrs. Aida Farha from the Saab Medical Library at AUBMC for assisting with the search strategy. We thank Ms. Lara Kahale for her contribution in obtaining the full text articles.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

Conceiving and designing the review: AH, RT, SK, EA. Coordinating the review: EA. Data extraction: AH, RT, RB, BD. Data analyses: EA. Data interpretation: SK, EA. Writing of the review: AH, EA. Reviewed and approved the final version of the manuscript: all authors.
Literatur
2.
Zurück zum Zitat Panlilio AL, Orelien JG, Srivastava PU, Jagger J, Cohn RD, Carco DM, et al. Estimate of the annual number of per- cutaneous injuries among hospital-based healthcare workers in the United States, 1997–1998. Infect Control Hosp Epidemiol. 2004;25(7):556–62.CrossRefPubMed Panlilio AL, Orelien JG, Srivastava PU, Jagger J, Cohn RD, Carco DM, et al. Estimate of the annual number of per- cutaneous injuries among hospital-based healthcare workers in the United States, 1997–1998. Infect Control Hosp Epidemiol. 2004;25(7):556–62.CrossRefPubMed
3.
Zurück zum Zitat National Audit Office. Office NA: 2003. National Audit Office. Office NA: 2003.
4.
Zurück zum Zitat Collins CH, Kennedy DA. Microbiological hazards of occupational needlestick and other sharps’ injuries. J Appl Bacteriol. 1987;62:385–402.CrossRefPubMed Collins CH, Kennedy DA. Microbiological hazards of occupational needlestick and other sharps’ injuries. J Appl Bacteriol. 1987;62:385–402.CrossRefPubMed
5.
Zurück zum Zitat Wagner D, de With K, Huzly D, Hufert F, Weidmann M, Breisinger S, et al. Nosocomial transmission of dengue. Emerg Infect Dis. 2004;10(10):1872–3.CrossRefPubMedPubMedCentral Wagner D, de With K, Huzly D, Hufert F, Weidmann M, Breisinger S, et al. Nosocomial transmission of dengue. Emerg Infect Dis. 2004;10(10):1872–3.CrossRefPubMedPubMedCentral
7.
Zurück zum Zitat Schmid GP, Buve A, Mugyenyi P, Garnett GP, Hayes RJ, Williams BG, et al. Transmission of HIV-1 infection in sub-Saharan Africa and effect of elimination of unsafe injections. Lancet. 2004;363:482–8.CrossRefPubMed Schmid GP, Buve A, Mugyenyi P, Garnett GP, Hayes RJ, Williams BG, et al. Transmission of HIV-1 infection in sub-Saharan Africa and effect of elimination of unsafe injections. Lancet. 2004;363:482–8.CrossRefPubMed
9.
Zurück zum Zitat Fisman DN, Mittleman MA, Sorock GS, Harris AD. Willingness to pay to avoid sharps-related injuries: a study in injured health care workers. Am J Infect Control. 2002;30(5):283–7.CrossRefPubMed Fisman DN, Mittleman MA, Sorock GS, Harris AD. Willingness to pay to avoid sharps-related injuries: a study in injured health care workers. Am J Infect Control. 2002;30(5):283–7.CrossRefPubMed
10.
Zurück zum Zitat Sohn JW, Kim BG, Kim SH, Han C. Mental health of healthcare workers who experience needlestick and sharps injuries. J Occup Health. 2006;48(6):474–9.CrossRefPubMed Sohn JW, Kim BG, Kim SH, Han C. Mental health of healthcare workers who experience needlestick and sharps injuries. J Occup Health. 2006;48(6):474–9.CrossRefPubMed
11.
Zurück zum Zitat American Nurses Association’s. Needlestick prevention guide. 2002. American Nurses Association’s. Needlestick prevention guide. 2002.
12.
Zurück zum Zitat World Health Organization. Quantifying selected major risks to health. In: The World Health Report. Geneva: WHO; 2002. p. 47–98. World Health Organization. Quantifying selected major risks to health. In: The World Health Report. Geneva: WHO; 2002. p. 47–98.
13.
Zurück zum Zitat Dziekan G, Chisholm D, Johns B, Rovira J, Hutin YJ. The cost-effectiveness of policies for the safe and appropriate use of injection in healthcare settings. Bull World Health Organ. 2003;81(4):277–85.PubMedPubMedCentral Dziekan G, Chisholm D, Johns B, Rovira J, Hutin YJ. The cost-effectiveness of policies for the safe and appropriate use of injection in healthcare settings. Bull World Health Organ. 2003;81(4):277–85.PubMedPubMedCentral
14.
Zurück zum Zitat Tuma S, Sepkowitz KA. Efficacy of safety-engineered device implementation in the prevention of percutaneous injuries: a review of published studies. Clin Infect Dis. 2006;42(8):1159–70. doi:10.1086/501456.CrossRefPubMed Tuma S, Sepkowitz KA. Efficacy of safety-engineered device implementation in the prevention of percutaneous injuries: a review of published studies. Clin Infect Dis. 2006;42(8):1159–70. doi:10.​1086/​501456.CrossRefPubMed
15.
Zurück zum Zitat Whitby M, McLaws M, Slater K. Needlestick injuries in a major teaching hospital: the worthwhile effect of hospital-wide replacement of conventional hollow-bore needles. Am J Infect Control. 2008;36(3):180–6.CrossRefPubMed Whitby M, McLaws M, Slater K. Needlestick injuries in a major teaching hospital: the worthwhile effect of hospital-wide replacement of conventional hollow-bore needles. Am J Infect Control. 2008;36(3):180–6.CrossRefPubMed
16.
Zurück zum Zitat Kontopantelis E, Reeves D. Performance of statistical methods for meta-analysis when true study effects are non-normally distributed: a simulation study. Stat Methods Med Res. 2012;21(4):409–26.CrossRefPubMed Kontopantelis E, Reeves D. Performance of statistical methods for meta-analysis when true study effects are non-normally distributed: a simulation study. Stat Methods Med Res. 2012;21(4):409–26.CrossRefPubMed
17.
Zurück zum Zitat Brockwell SE, Gordon IR. A comparison of statistical methods for meta-analysis. Stat Med. 2001;20(6):825–40.CrossRefPubMed Brockwell SE, Gordon IR. A comparison of statistical methods for meta-analysis. Stat Med. 2001;20(6):825–40.CrossRefPubMed
18.
Zurück zum Zitat Reddy SG, Emery RJ. Assessing the effect of long-term availability of engineering controls on needlestick injuries among health care workers: a 3-year preimplementation and postimplementation comparison. Am J Infect Control. 2001;2(6):425–7.CrossRef Reddy SG, Emery RJ. Assessing the effect of long-term availability of engineering controls on needlestick injuries among health care workers: a 3-year preimplementation and postimplementation comparison. Am J Infect Control. 2001;2(6):425–7.CrossRef
19.
Zurück zum Zitat Adams D, Elliott TS. Impact of safety needle devices on occupationally acquired needlestick injuries: a four-year prospective study. J Hosp Infect. 2006;64(1):50–5.CrossRefPubMed Adams D, Elliott TS. Impact of safety needle devices on occupationally acquired needlestick injuries: a four-year prospective study. J Hosp Infect. 2006;64(1):50–5.CrossRefPubMed
20.
21.
Zurück zum Zitat Younger B, Hnt EH, Robinson C, McLemore C. Impact of a shielded safety syringe on needlestick injuries among healthcare workers. Infect Control Hosp Epidemiol. 1992;13(6):349–53.CrossRefPubMed Younger B, Hnt EH, Robinson C, McLemore C. Impact of a shielded safety syringe on needlestick injuries among healthcare workers. Infect Control Hosp Epidemiol. 1992;13(6):349–53.CrossRefPubMed
22.
Zurück zum Zitat Duesman K, Ross J. Survey of accidental needlesticks in 26 facilities using Vanishpoint automated retraction syringe. J Healthc Saf Compliance Infect Control. 1998;2(3):111–4. Duesman K, Ross J. Survey of accidental needlesticks in 26 facilities using Vanishpoint automated retraction syringe. J Healthc Saf Compliance Infect Control. 1998;2(3):111–4.
23.
Zurück zum Zitat Guyatt G, Oxman AD, Oxman AD, Akl EA, Kunz R, Vist G, et al. GRADE guidelines: 1. Introduction-GRADE evidence profiles and summary of findings tables. J Clin Epidemiol. 2011;64(4):383–94.CrossRefPubMed Guyatt G, Oxman AD, Oxman AD, Akl EA, Kunz R, Vist G, et al. GRADE guidelines: 1. Introduction-GRADE evidence profiles and summary of findings tables. J Clin Epidemiol. 2011;64(4):383–94.CrossRefPubMed
24.
Zurück zum Zitat Sohn S, Eagan J, Sepkowitz KA, Zccotti G. Effect of implementing safety-engineered devices on percutaneous injury epidemiology. Infect Control Hosp Epidemiol. 2004;25(7):536–42.CrossRefPubMed Sohn S, Eagan J, Sepkowitz KA, Zccotti G. Effect of implementing safety-engineered devices on percutaneous injury epidemiology. Infect Control Hosp Epidemiol. 2004;25(7):536–42.CrossRefPubMed
25.
Zurück zum Zitat van der Molen HF, Zwinderman KAH, Sluiter JK, Frings-Dresen MHW. Better effect of the use of a needle safety device in combination with an interactive workshop to prevent needle stick injuries. Saf Sci. 2011;4(8–9):1180–6.CrossRef van der Molen HF, Zwinderman KAH, Sluiter JK, Frings-Dresen MHW. Better effect of the use of a needle safety device in combination with an interactive workshop to prevent needle stick injuries. Saf Sci. 2011;4(8–9):1180–6.CrossRef
26.
Zurück zum Zitat Hoffmann C, Bchholz L, Schnitzler P. Reduction of needlestick injuries in healthcare personnel at a university hospital using safety devices. J Occup Med Toxicol. 2013;8(1):20.CrossRefPubMedPubMedCentral Hoffmann C, Bchholz L, Schnitzler P. Reduction of needlestick injuries in healthcare personnel at a university hospital using safety devices. J Occup Med Toxicol. 2013;8(1):20.CrossRefPubMedPubMedCentral
27.
Zurück zum Zitat Valls V, Lozano MS, Yanez R, Martinez MJ, Pascal F, Lloret J, et al. Use of safety devices and the prevention of percutaneous injuries among healthcare workers (Provisional abstract). Infect Control Hosp Epidemiol. 2007;28:1352–60.CrossRefPubMed Valls V, Lozano MS, Yanez R, Martinez MJ, Pascal F, Lloret J, et al. Use of safety devices and the prevention of percutaneous injuries among healthcare workers (Provisional abstract). Infect Control Hosp Epidemiol. 2007;28:1352–60.CrossRefPubMed
28.
Zurück zum Zitat Adams D, Elliott TS. A comparative user evaluation of three needle-protective devices. Br J Nurs (Mark Allen Publishing). 2003;12(8):470–4.CrossRef Adams D, Elliott TS. A comparative user evaluation of three needle-protective devices. Br J Nurs (Mark Allen Publishing). 2003;12(8):470–4.CrossRef
29.
Zurück zum Zitat Azar-Cavanagh M, Burdt P, Green-McKenzie J. Effect of the introduction of an engineered sharps injury prevention device on the percutaneous injury rate in healthcare workers. Infect Control Hosp Epidemiol. 2007;28(2):165–70.CrossRefPubMed Azar-Cavanagh M, Burdt P, Green-McKenzie J. Effect of the introduction of an engineered sharps injury prevention device on the percutaneous injury rate in healthcare workers. Infect Control Hosp Epidemiol. 2007;28(2):165–70.CrossRefPubMed
30.
Zurück zum Zitat Beason R, Borgignon J, Fowler D, Gardner C. Evaluation of a needle-free intravenous access system. J Intraven Nurs. 1992;15(1):11–6.PubMed Beason R, Borgignon J, Fowler D, Gardner C. Evaluation of a needle-free intravenous access system. J Intraven Nurs. 1992;15(1):11–6.PubMed
31.
Zurück zum Zitat Evaluation of blunt suture needles in preventing percutaneous injuries among health-care workers during gynecologic surgical procedures -- New York City, March 1993-June 1994. MMWR Morb Mortal Wkly Rep. 1997;46(2):25–29. Evaluation of blunt suture needles in preventing percutaneous injuries among health-care workers during gynecologic surgical procedures -- New York City, March 1993-June 1994. MMWR Morb Mortal Wkly Rep. 1997;46(2):25–29.
32.
Zurück zum Zitat Armadans Gil L, Fernandez Cano MI, Albero Andres I, Angles Mellado ML, Sanchez Garcia JM, Campins Marti M, et al. Safety-engineered devices to prevent percutaneous injuries: cost-effectiveness analysis on prevention of high-risk exposure. [Spanish] Analisis coste-efectividad de dispositivos sanitarios disenados para prevenir exposiciones percutaneas. Gac Sanit. 2006;20(5):374–81.CrossRefPubMed Armadans Gil L, Fernandez Cano MI, Albero Andres I, Angles Mellado ML, Sanchez Garcia JM, Campins Marti M, et al. Safety-engineered devices to prevent percutaneous injuries: cost-effectiveness analysis on prevention of high-risk exposure. [Spanish] Analisis coste-efectividad de dispositivos sanitarios disenados para prevenir exposiciones percutaneas. Gac Sanit. 2006;20(5):374–81.CrossRefPubMed
33.
Zurück zum Zitat Billiet LS, Parker CR, Tanley PC, Wallas CH. Needlestick injury rate reduction during phlebotomy: a comparative study of two safety devices. Lab Med. 1991;22(2):120–3.CrossRef Billiet LS, Parker CR, Tanley PC, Wallas CH. Needlestick injury rate reduction during phlebotomy: a comparative study of two safety devices. Lab Med. 1991;22(2):120–3.CrossRef
34.
Zurück zum Zitat Catalán Gómez MT, Sol Vidiella J, Castellà M, Castells Bo C, Losada Pla N, Llís Espny J. Implementation of safety devices: biological accident prevention [Spanish]. Rev Enferm. 2010;3(4):50–4. Catalán Gómez MT, Sol Vidiella J, Castellà M, Castells Bo C, Losada Pla N, Llís Espny J. Implementation of safety devices: biological accident prevention [Spanish]. Rev Enferm. 2010;3(4):50–4.
35.
Zurück zum Zitat Dugger B. Introducing products to prevent needlesticks. Nurs Manag. 1992;23(10):62–6.CrossRef Dugger B. Introducing products to prevent needlesticks. Nurs Manag. 1992;23(10):62–6.CrossRef
36.
Zurück zum Zitat Edwards C, Johnson C. Evaluation of a luer-activated intravenous administration system. J Assoc Vasc Access. 2012;17(4):200–7.CrossRef Edwards C, Johnson C. Evaluation of a luer-activated intravenous administration system. J Assoc Vasc Access. 2012;17(4):200–7.CrossRef
37.
Zurück zum Zitat Gartner K. Impact of a needleless intravenous system in a university Hospital. Am J Infect Control. 1992;20(2):75–9.CrossRefPubMed Gartner K. Impact of a needleless intravenous system in a university Hospital. Am J Infect Control. 1992;20(2):75–9.CrossRefPubMed
38.
Zurück zum Zitat Gershon RRM, Pearse L, Grimes M, Flanagan PA, Vlahov D. The impact of multifocused interventions on sharps injury rates at an acute-care hospital. Infect Control Hosp Epidemiol. 1999;20(12):806–11.CrossRefPubMed Gershon RRM, Pearse L, Grimes M, Flanagan PA, Vlahov D. The impact of multifocused interventions on sharps injury rates at an acute-care hospital. Infect Control Hosp Epidemiol. 1999;20(12):806–11.CrossRefPubMed
39.
Zurück zum Zitat Lamontagne F, Abitebol D, Lolom I, Pellissier G, Tarantola A, Descamps JM, et al. Role of safety-engineered devices in preventing needlestick injuries in 32 French hospitals. Infect Control Hosp Epidemiol. 2007;28(1):18–23.CrossRefPubMed Lamontagne F, Abitebol D, Lolom I, Pellissier G, Tarantola A, Descamps JM, et al. Role of safety-engineered devices in preventing needlestick injuries in 32 French hospitals. Infect Control Hosp Epidemiol. 2007;28(1):18–23.CrossRefPubMed
40.
Zurück zum Zitat Laufer FN, Chiarello LA. Application of cost-effectiveness methodology to the consideration of needlestick-prevention technology. Am J Infect Control. 1994;22(2):75–82.CrossRefPubMed Laufer FN, Chiarello LA. Application of cost-effectiveness methodology to the consideration of needlestick-prevention technology. Am J Infect Control. 1994;22(2):75–82.CrossRefPubMed
41.
Zurück zum Zitat Lawrence LW, Delclos GL, Felknor SA, Johnson PC, Frankowski RF, Cooper SP, et al. The effectiveness of a needleless intravenous connection system: an assessment by injury rate and user satisfaction. Infect Control Hosp Epidemiol. 1997;18(3):175–82.CrossRefPubMed Lawrence LW, Delclos GL, Felknor SA, Johnson PC, Frankowski RF, Cooper SP, et al. The effectiveness of a needleless intravenous connection system: an assessment by injury rate and user satisfaction. Infect Control Hosp Epidemiol. 1997;18(3):175–82.CrossRefPubMed
42.
Zurück zum Zitat L’Ecyer PB, Schwab EO, Iademarco E, Barr N, Aton EA, Fraser VJ. Randomized prospective study of the impact of three needleless intravenous systems on needlestick injury rates. Infect Control Hosp Epidemiol. 1996;17(12):803–8.CrossRef L’Ecyer PB, Schwab EO, Iademarco E, Barr N, Aton EA, Fraser VJ. Randomized prospective study of the impact of three needleless intravenous systems on needlestick injury rates. Infect Control Hosp Epidemiol. 1996;17(12):803–8.CrossRef
43.
Zurück zum Zitat MacPherson J. The interlink needleless intravenous system did not reduce the number of needlestick injuries in Christchurch hospital operating theatres. N Z Med J. 1996;10(1031):387–8. MacPherson J. The interlink needleless intravenous system did not reduce the number of needlestick injuries in Christchurch hospital operating theatres. N Z Med J. 1996;10(1031):387–8.
44.
Zurück zum Zitat McCleary J, Caldero K, Adams T. Guarded fistula needle reduces needlestick injuries in hemodialysis. Nephrol News Issues. 2002;16(6):66–70. 72.PubMed McCleary J, Caldero K, Adams T. Guarded fistula needle reduces needlestick injuries in hemodialysis. Nephrol News Issues. 2002;16(6):66–70. 72.PubMed
45.
Zurück zum Zitat Mendelson MH, Short LJ, Schechter CB, Meyers BR, Rodrigez M, Cohen S, et al. Study of a needleless intermittent intravenous-access system for peripheral infusions: analysis of staff, patient, and institutional outcomes. Infect Control Hosp Epidemiol. 1998;1(6):401–6.CrossRef Mendelson MH, Short LJ, Schechter CB, Meyers BR, Rodrigez M, Cohen S, et al. Study of a needleless intermittent intravenous-access system for peripheral infusions: analysis of staff, patient, and institutional outcomes. Infect Control Hosp Epidemiol. 1998;1(6):401–6.CrossRef
46.
Zurück zum Zitat Mlherin S, Rickman LS, Jackson MM. Initial worker evaluation of a new safety syringe. Infect Control Hosp Epidemiol. 1996;17(9):593–4.CrossRef Mlherin S, Rickman LS, Jackson MM. Initial worker evaluation of a new safety syringe. Infect Control Hosp Epidemiol. 1996;17(9):593–4.CrossRef
47.
Zurück zum Zitat Orenstein R, Reynolds L, Karabaic M, Lamb A, Markowitz SM, Wong ES. Do protective devices prevent needlestick injuries among health care workers? Am J Infect Control. 1995;23(6):344–51.CrossRefPubMed Orenstein R, Reynolds L, Karabaic M, Lamb A, Markowitz SM, Wong ES. Do protective devices prevent needlestick injuries among health care workers? Am J Infect Control. 1995;23(6):344–51.CrossRefPubMed
48.
Zurück zum Zitat Pereira CC, Bishai D. Vaccine presentation in the USA: economics of prefilled syringes versus multidose vials for influenza vaccination. Expert Rev Vaccines. 2010;11:1343–9.CrossRef Pereira CC, Bishai D. Vaccine presentation in the USA: economics of prefilled syringes versus multidose vials for influenza vaccination. Expert Rev Vaccines. 2010;11:1343–9.CrossRef
49.
Zurück zum Zitat Pro VDC, Segata G, Piccini A, Argentero G, Signorini PA, Daglio L, et al. Theme updates on the epidemiology of occupational bloodborne infectious diseases. [Italian] Aggiornamenti in tema di epidemiologia delle malattie infettive occupazionali trasmesse per via ematica. G Ital Med Lav Ergon. 2010;32(3):235–9. Pro VDC, Segata G, Piccini A, Argentero G, Signorini PA, Daglio L, et al. Theme updates on the epidemiology of occupational bloodborne infectious diseases. [Italian] Aggiornamenti in tema di epidemiologia delle malattie infettive occupazionali trasmesse per via ematica. G Ital Med Lav Ergon. 2010;32(3):235–9.
50.
Zurück zum Zitat Roges A, Verdn-Esqer C, Buisson-Valles I, Laville M, Lashéras A, Sarrat A, et al. Impact of safety devices for preventing percutaneous injuries related to phlebotomy procedures in health care workers. Am J Infect Control. 2004;32(8):441–4.CrossRef Roges A, Verdn-Esqer C, Buisson-Valles I, Laville M, Lashéras A, Sarrat A, et al. Impact of safety devices for preventing percutaneous injuries related to phlebotomy procedures in health care workers. Am J Infect Control. 2004;32(8):441–4.CrossRef
51.
Zurück zum Zitat Skolnick R, LaRocca J, Barba D, Paicius L. Evaluation and implementation of a needleless intravenous system: making needlesticks a needless problem. Am J Infect Control. 1993;21(1):39–41.CrossRefPubMed Skolnick R, LaRocca J, Barba D, Paicius L. Evaluation and implementation of a needleless intravenous system: making needlesticks a needless problem. Am J Infect Control. 1993;21(1):39–41.CrossRefPubMed
52.
Zurück zum Zitat Terrell F, Williams B. Implementation of a customized needleless intravenous delivery system. J Intraven Nurs. 1993;16(6):339–44.PubMed Terrell F, Williams B. Implementation of a customized needleless intravenous delivery system. J Intraven Nurs. 1993;16(6):339–44.PubMed
53.
Zurück zum Zitat Vadelle-Malbos C, Gury C, Prost G, Brossard D, V-Thi P, Bonaccorsi A. Sterile needleprick prevention devices: assessment after recording of accidents in two hospitals. [Italian] Dispositifs medicaux steriles de protection des piqures: evaluation apres recensement des accidents dans deux hopitaux. Archives des Maladies Professionnelles et de Medecine du Travail. 1996;57(7):508–18. Vadelle-Malbos C, Gury C, Prost G, Brossard D, V-Thi P, Bonaccorsi A. Sterile needleprick prevention devices: assessment after recording of accidents in two hospitals. [Italian] Dispositifs medicaux steriles de protection des piqures: evaluation apres recensement des accidents dans deux hopitaux. Archives des Maladies Professionnelles et de Medecine du Travail. 1996;57(7):508–18.
54.
Zurück zum Zitat Wolfrm J. A follow-up evaluation to a needle-free I.V. system. Nurs Manag. 1994;25(12):33–5.CrossRef Wolfrm J. A follow-up evaluation to a needle-free I.V. system. Nurs Manag. 1994;25(12):33–5.CrossRef
55.
Zurück zum Zitat Yassi A, McGill ML, Khokhar JB. Efficacy and cost-effectiveness of a needleless intravenous access system. Am J Infect Control. 1995;23(2):57–64.CrossRefPubMed Yassi A, McGill ML, Khokhar JB. Efficacy and cost-effectiveness of a needleless intravenous access system. Am J Infect Control. 1995;23(2):57–64.CrossRefPubMed
56.
Zurück zum Zitat Zakrzewska JM, Greenwood I, Jackson J. Cross-infection control: introducing safety syringes into a UK dental school- a controlled study. Br Dent J. 2001;190(2):88–92.PubMed Zakrzewska JM, Greenwood I, Jackson J. Cross-infection control: introducing safety syringes into a UK dental school- a controlled study. Br Dent J. 2001;190(2):88–92.PubMed
57.
Zurück zum Zitat Peate WF. Preventing needlesticks in emergency medical system workers. J Occup Environ Med. 2001;43(6):554–7.CrossRefPubMed Peate WF. Preventing needlesticks in emergency medical system workers. J Occup Environ Med. 2001;43(6):554–7.CrossRefPubMed
58.
Zurück zum Zitat Valls V, Lozano MS, Yanez R, Martinez MJ, Pascal F, Lloret J, et al. Use of safety devices and the prevention of percutaneous injuries among healthcare workers [corrected] [published erratum appears in INFECT CONTROL HOSP EPIDEMIOL 2008 Mar;29(3):288]. Infect Control Hosp Epidemiol. 2007;28(12):1352–60.CrossRefPubMed Valls V, Lozano MS, Yanez R, Martinez MJ, Pascal F, Lloret J, et al. Use of safety devices and the prevention of percutaneous injuries among healthcare workers [corrected] [published erratum appears in INFECT CONTROL HOSP EPIDEMIOL 2008 Mar;29(3):288]. Infect Control Hosp Epidemiol. 2007;28(12):1352–60.CrossRefPubMed
59.
Zurück zum Zitat Kontopantelis E, Springate DA, Reeves D. A re-analysis of the Cochrane Library data: the dangers of unobserved heterogeneity in meta-analyses. PLoS One. 2013;8(7):e69930.CrossRefPubMedPubMedCentral Kontopantelis E, Springate DA, Reeves D. A re-analysis of the Cochrane Library data: the dangers of unobserved heterogeneity in meta-analyses. PLoS One. 2013;8(7):e69930.CrossRefPubMedPubMedCentral
60.
Zurück zum Zitat Kicinski M, Springate DA, Kontopantelis E. Publication bias in meta-analyses from the Cochrane Database of Systematic Reviews. Stat Med. 2015;34(20):2781–93.CrossRefPubMed Kicinski M, Springate DA, Kontopantelis E. Publication bias in meta-analyses from the Cochrane Database of Systematic Reviews. Stat Med. 2015;34(20):2781–93.CrossRefPubMed
61.
Zurück zum Zitat Beswick A, Robinson E, Evans G, Codling A. An evaluation of the efficacy of safer sharps devices. Systematic review. Prepared by the Health and Safety Laboratory for the Health and Safety Executive 2012. 2012. Beswick A, Robinson E, Evans G, Codling A. An evaluation of the efficacy of safer sharps devices. Systematic review. Prepared by the Health and Safety Laboratory for the Health and Safety Executive 2012. 2012.
62.
Zurück zum Zitat Zakrzewska JM, Greenwood I, Jackson J. Introducing safety syringes into a UK dental school: a controlled study (Structured abstract). Br Dent J. 2001;190:88–92.PubMed Zakrzewska JM, Greenwood I, Jackson J. Introducing safety syringes into a UK dental school: a controlled study (Structured abstract). Br Dent J. 2001;190:88–92.PubMed
63.
Zurück zum Zitat Mustafa RA, Santesso N, Brozek J, Akl EA, Walter SD, Norman G, et al. The GRADE approach is reproducible in assessing the quality of evidence of quantitative evidence syntheses. J Clin Epidemiol. 2013;66(7):736–42. qiz 742 e731-735.CrossRefPubMed Mustafa RA, Santesso N, Brozek J, Akl EA, Walter SD, Norman G, et al. The GRADE approach is reproducible in assessing the quality of evidence of quantitative evidence syntheses. J Clin Epidemiol. 2013;66(7):736–42. qiz 742 e731-735.CrossRefPubMed
Metadaten
Titel
Safety engineered injection devices for intramuscular, subcutaneous and intradermal injections in healthcare delivery settings: a systematic review and meta-analysis
verfasst von
Alain C. Harb
Rami Tarabay
Batoul Diab
Rami A. Ballout
Selma Khamassi
Elie A. Akl
Publikationsdatum
01.12.2015
Verlag
BioMed Central
Erschienen in
BMC Nursing / Ausgabe 1/2015
Elektronische ISSN: 1472-6955
DOI
https://doi.org/10.1186/s12912-015-0119-1